High Precision Devices | 303.447.2558

Cryostat Model 102 Denali

Small, rectangular, pulse-tube ADR cryostat

Features:

  • Cryomech PT407 Pulse Tube Cooler
  • Remote Rotary Valve Motor
  • Vibration isolation bellows
  • HPD ADR (NIST design)
    • Kevlar Suspension
    • 1 K & 50 mK Stages
  • Rectangular cross section for close positioning to a scanning electron microscope.
  • Robust internal construction
  • Custom Service Stand (Option)
    • Allows for transport & servicing
    • 30 inches of height adjustment

Specifications:

  • Experimental Space: 5.5" X 9.5" X 6"
  • Cool down from 300 K to 4 K in 12 hours
  • Pulse-tube stage temperatures of 50 K and 2.7 K
  • 700 mW cooling power at 4.2 K
  • ADR base temperature of 50 mK
  • >150 hour no-load regulation at 100 mK
102 Denali ADR Cryostat, Cryocooler

Section view of 102 Denali ADR cryostat showing pulse-tube head, adiabatic demagnetazition refrigerator, high temp superconducting leads for 4T magnet, thermal shielding, mechanical heat switch.


102 Denali ADR Cryostat, Cryocooler

Interior view of 102 Denali ADR cryostat on a service stand, showing pulse-tube head, thermal links at 50K and 3K stages, detector mounts, vibration isolation bellows.

102 Denali ADR Cryostat, Cryocooler

Model 102 Denali ADR cryostat on Atlas 100 service stand.

102 Denali cryostat

Model 102 Denali ADR cryostat on Atlas 100 service stand.

102 Denali cryostat

Cross-section of 102 Denali cryostat showing pulse-tube cooler, ADR stage, and thermal shields.

102 Denali cryostat

Top view of 102 Denali cryostat showing pulse tube cooler and feedthroughs.


102 Denali ADR Cryostat, Cryocooler

Typical cooling cycle for pulse-tube cooled ADR cryostat.


Compare Models

Wet DR Cryostat
Pulse Tube ADR Cryostat
Pulse Tube ADR Cryostat
Pulse Tube ADR Cryostat
Dry DR Cryostat
Pulse Tube ADR Cryostat
Vacuum Jacket Size
Ø 54 cm X 119 cm Tall
Vacuum Jacket Size
33 cm X 22 cm X
66 cm Tall
Vacuum Jacket Size
Ø 35 cm X 69 cm Tall
Vacuum Jacket Size
Ø 55 cm X 110 cm Tall
Vacuum Jacket Size
Ø 55 cm X 110 cm Tall
Vacuum Jacket Size
Ø 44 cm X 65 cm Tall
Experimental Volume
Ø 42 cm X 19 cm Tall
Experimental Volume
24 cm X 15 cm X
14 cm Tall
Experimental Volume
Ø 26 cm X 25 cm Tall
Experimental Volume
Ø 44 cm X 60 cm Tall
Experimental Volume
Ø 38 cm X 19 cm Tall
Experimental Volume
Ø 34 cm X 21 cm Tall
1st Stage Cooling Power
77 K LN2
1st Stage Cooling Power
25 W @ 55 K
1st Stage Cooling Power
25 W @ 55 K
1st Stage Cooling Power
35 W @ 45 K
40 W @ 45 K
1st Stage Cooling Power
40 W @ 45 K
1st Stage Cooling Power
25 W @ 55 K
35 W @ 45 K
40 W @ 45 K
2nd Stage Cooling Power
4K LHe
2nd Stage Cooling Power
0.7 W @ 4.2 K
2nd Stage Cooling Power
0.7 W @ 4.2 K
2nd Stage Cooling Power
1 W @ 4.2 K
1.5 W @ 4.2 K
2nd Stage Cooling Power
1.5 W @ 4.2 K
2nd Stage Cooling Power
0.7 W @ 4.2 K
1 W @ 4.2 K
1.5 W @ 4.2 K
Lower Stages
1 K LHe Pot
.6 K Still
.1 K Cold Plate
GGG Cooling Capacity
1.2 J @ 1 K
GGG Cooling Capacity
1.2 J @ 1 K
GGG Cooling Capacity
1.2 J @ 1 K
Lower Stages
1.5 K
.6 K Still
.1 K Cold Plate
GGG Cooling Capacity
1.2 J @ 1 K
Mixing Chamber Base Temperature
30 mK
ADR Base Temperature
< 50 mK
ADR Base Temperature
< 50 mK
ADR Base Temperature
< 50 mK
Mixing Chamber Base Temperature
30 mK
ADR Base Temperature
< 50 mK
Mixing Chamber Cooling Power
150 µW @ 100 mK
FAA Cooling Capacity
118 mJ @ 100 mK
FAA Cooling Capacity
118 mJ @ 100 mK
FAA Cooling Capacity
118 mJ @ 100 mK
Mixing Chamber Cooling Power
200 µW @ 100 mK
FAA Cooling Capacity
118 mJ @ 100 mK