High Precision Devices | 303.447.2558

Cryostat Model 103 Rainier

Small, cylindrical, pulse-tube ADR cryostat

Features:

  • Cryomech PT407 Pulse Tube Cooler
  • Remote Rotary Valve Motor
  • Vibration isolation bellows
  • Electronically controlled motorized heat switch
  • Quick release flanges on vacuum jackets
  • Nickel plated aluminum thermal shields (50 K, 4 K)
  • HPD ADR (NIST design)
    • Kevlar Suspension
    • 1 K & 50 mK Stages
  • Optimized for versatility and flexibility
    • 4X NW50 ports on top flange
    • 2X NW25 ports on top flange
    • 1X NW63 port on vacuum jacket side
    • 1X Ø6" port on vacuum jacket bottom
  • Robust internal construction
  • Custom Service Stand (Option)
    • Allows for transport & servicing
    • 30 inches of height adjustment

Specifications:

  • Large Experimental Space: 10.2" Diameter X 10" Tall
  • Cool down from 300 K to 4 K in 14 hours
  • Pulse-tube stage temperatures of 50 K and 2.7 K
  • 700 mW cooling power at 4.2 K
  • ADR base temperature of 50 mK
  • >200 hour no-load regulation at 100 mK
103 Rainier ADR Cryostat, Cryocooler

Section view of 103 Rainier ADR cryostat showing pulse-tube head, adiabatic demagnetization refrigerator, high temp superconducting leads for 4 T magnet, thermal shielding.


103 Rainier ADR Cryostat, Cryocooler

Model 103 Rainier ADR cryostat showing vacuum jacket, side port, pulse-tube head and anti-vibration bellows.

103 Rainier ADR Cryostat, Cryocooler

Model 103 Rainier ADR cryostat showing
on the Model 100 Atlas service stand.

103 Rainier ADR Cryostat, Cryocooler

Side view of Model 103 Rainier on the Model 100 Atlas service stand.

103 Rainier cryostat

Side view of 103 Rainier ADR cryostat vacuum jacket, side port, pulse-tube head.

103 Rainier cryostat

Cross-section of 103 Rainier cryostat showing pulse-tube cooler, ADR, thermal links and shields.

103 Rainier cryostat

Top view of 103 Rainier cryostat showing pulse-tube cooler and feedthrough flanges.


103 Rainier ADR Cryostat, Cryocooler

Typical cooling cycle for pulse-tube cooled ADR cryostat.


Compare Models

Wet DR Cryostat
Pulse Tube ADR Cryostat
Pulse Tube ADR Cryostat
Pulse Tube ADR Cryostat
Dry DR Cryostat
Pulse Tube ADR Cryostat
Vacuum Jacket Size
Ø 54 cm X 119 cm Tall
Vacuum Jacket Size
33 cm X 22 cm X
66 cm Tall
Vacuum Jacket Size
Ø 35 cm X 69 cm Tall
Vacuum Jacket Size
Ø 55 cm X 110 cm Tall
Vacuum Jacket Size
Ø 55 cm X 110 cm Tall
Vacuum Jacket Size
Ø 44 cm X 65 cm Tall
Experimental Volume
Ø 42 cm X 19 cm Tall
Experimental Volume
24 cm X 15 cm X
14 cm Tall
Experimental Volume
Ø 26 cm X 25 cm Tall
Experimental Volume
Ø 44 cm X 60 cm Tall
Experimental Volume
Ø 38 cm X 19 cm Tall
Experimental Volume
Ø 34 cm X 21 cm Tall
1st Stage Cooling Power
77 K LN2
1st Stage Cooling Power
25 W @ 55 K
1st Stage Cooling Power
25 W @ 55 K
1st Stage Cooling Power
35 W @ 45 K
40 W @ 45 K
1st Stage Cooling Power
40 W @ 45 K
1st Stage Cooling Power
25 W @ 55 K
35 W @ 45 K
40 W @ 45 K
2nd Stage Cooling Power
4K LHe
2nd Stage Cooling Power
0.7 W @ 4.2 K
2nd Stage Cooling Power
0.7 W @ 4.2 K
2nd Stage Cooling Power
1 W @ 4.2 K
1.5 W @ 4.2 K
2nd Stage Cooling Power
1.5 W @ 4.2 K
2nd Stage Cooling Power
0.7 W @ 4.2 K
1 W @ 4.2 K
1.5 W @ 4.2 K
Lower Stages
1 K LHe Pot
.6 K Still
.1 K Cold Plate
GGG Cooling Capacity
1.2 J @ 1 K
GGG Cooling Capacity
1.2 J @ 1 K
GGG Cooling Capacity
1.2 J @ 1 K
Lower Stages
1.5 K
.6 K Still
.1 K Cold Plate
GGG Cooling Capacity
1.2 J @ 1 K
Mixing Chamber Base Temperature
30 mK
ADR Base Temperature
< 50 mK
ADR Base Temperature
< 50 mK
ADR Base Temperature
< 50 mK
Mixing Chamber Base Temperature
30 mK
ADR Base Temperature
< 50 mK
Mixing Chamber Cooling Power
150 µW @ 100 mK
FAA Cooling Capacity
118 mJ @ 100 mK
FAA Cooling Capacity
118 mJ @ 100 mK
FAA Cooling Capacity
118 mJ @ 100 mK
Mixing Chamber Cooling Power
200 µW @ 100 mK
FAA Cooling Capacity
118 mJ @ 100 mK