High Precision Devices | 303.447.2558

HPD Research Cryostats

Introducing our model 103 Rapid Cool Cryostat
< 4 hours to < 4K!

Select from one of our current cryostat models or re-configure one to your needs. These precision, pulse-tube coolers are designed for a range of scientific applications, including infrared spectroscopy and alpha particle detection.

101 Cryostat 102 Cryostat 103 Cryostat 104 Cryostat 105 Cryostat 106 Cryostat
Wet DR Cryostat
Pulse Tube ADR Cryostat
Pulse Tube ADR Cryostat
Pulse Tube ADR Cryostat
Dry DR Cryostat
Pulse Tube ADR Cryostat
Vacuum Jacket Size
Ø 54 cm X 119 cm Tall
Vacuum Jacket Size
33 cm X 22 cm X
66 cm Tall
Vacuum Jacket Size
Ø 35 cm X 69 cm Tall
Vacuum Jacket Size
Ø 55 cm X 110 cm Tall
Vacuum Jacket Size
Ø 55 cm X 110 cm Tall
Vacuum Jacket Size
Ø 44 cm X 65 cm Tall
Experimental Volume
Ø 42 cm X 19 cm Tall
Experimental Volume
24 cm X 15 cm X
14 cm Tall
Experimental Volume
Ø 26 cm X 25 cm Tall
Experimental Volume
Ø 44 cm X 60 cm Tall
Experimental Volume
Ø 38 cm X 19 cm Tall
Experimental Volume
Ø 34 cm X 21 cm Tall
1st Stage Cooling Power
77 K LN2
1st Stage Cooling Power
25 W @ 55 K
1st Stage Cooling Power
25 W @ 55 K
1st Stage Cooling Power
35 W @ 45 K
40 W @ 45 K
1st Stage Cooling Power
40 W @ 45 K
1st Stage Cooling Power
25 W @ 55 K
35 W @ 45 K
40 W @ 45 K
2nd Stage Cooling Power
4K LHe
2nd Stage Cooling Power
0.7 W @ 4.2 K
2nd Stage Cooling Power
0.7 W @ 4.2 K
2nd Stage Cooling Power
1 W @ 4.2 K
1.5 W @ 4.2 K
2nd Stage Cooling Power
1.5 W @ 4.2 K
2nd Stage Cooling Power
0.7 W @ 4.2 K
1 W @ 4.2 K
1.5 W @ 4.2 K
Lower Stages
1 K LHe Pot
.6 K Still
.1 K Cold Plate
GGG Cooling Capacity
1.2 J @ 1 K
GGG Cooling Capacity
1.2 J @ 1 K
GGG Cooling Capacity
1.2 J @ 1 K
Lower Stages
1.5 K
.6 K Still
.1 K Cold Plate
GGG Cooling Capacity
1.2 J @ 1 K
Mixing Chamber Base Temperature
30 mK
ADR Base Temperature
< 50 mK
ADR Base Temperature
< 50 mK
ADR Base Temperature
< 50 mK
Mixing Chamber Base Temperature
30 mK
ADR Base Temperature
< 50 mK
Mixing Chamber Cooling Power
150 µW @ 100 mK
FAA Cooling Capacity
118 mJ @ 100 mK
FAA Cooling Capacity
118 mJ @ 100 mK
FAA Cooling Capacity
118 mJ @ 100 mK
Mixing Chamber Cooling Power
200 µW @ 100 mK
FAA Cooling Capacity
118 mJ @ 100 mK